Decision 2426/QD-BYT 2021 Guidance for preparation of ASEAN common submission dossiers

  • Summary
  • Content
  • Status
  • Vietnamese
  • Download
Save

Please log in to use this function

Send link to email

Please log in to use this function

Error message
Font size:

ATTRIBUTE

Decision No. 2426/QD-BYT dated May 15, 2021 of the Ministry of Health on promulgation of Guidance for preparation of ASEAN common submission dossiers
Issuing body: Ministry of HealthEffective date:
Known

Please log in to a subscriber account to use this function.

Don’t have an account? Register here

Official number:2426/QD-BYTSigner:Truong Quoc Cuong
Type:DecisionExpiry date:Updating
Issuing date:15/05/2021Effect status:
Known

Please log in to a subscriber account to use this function.

Don’t have an account? Register here

Fields:Medical - Health
For more details, click here.
Download files here.
LuatVietnam.vn is the SOLE distributor of English translations of Official Gazette published by the Vietnam News Agency
Effect status: Known

THE MINISTRY OF HEALTH

_______

No. 2426/QD-BYT

THE SOCIALIST REPUBLIC OF VIETNAM

Independence - Freedom - Happiness

________________________

Hanoi, May 15, 2021

DECISION

On promulgation of Guidance for preparation of ASEAN common submission dossiers

___________

THE MINISTER OF HEALTH

 

Pursuant to the Government’s Decree No. 75/2017/ND-CP dated June 20, 2017, defining the functions, tasks, powers and organizational structure of the Ministry of Health;

Pursuant to the Government's Decree No. 36/2016/ND-CP dated May 15, 2016, on medical equipment management;

Pursuant to the Government’s Decree No. 169/2018/ND-CP dated December 31, 2018, on amending and supplementing a number of articles of the Government’s Decree No. 36/2016/ND-CP dated May 15, 2016, on medical equipment management;

Pursuant to the Government’s Decree No. 03/2020/ND-CP dated January 01, 2020 on amending and supplementing Article 68 of the Government’s Decree No. 36/2016/ND-CP dated May 25, 2016, on medical equipment management, which was amended and supplemented under the Government’s Decree No. 169/2018/ND-CP dated December 31, 2018, on amending and supplementing a number of articles of the Government's Decree No. 36/2016/ND-CP dated May 15, 2016, on medical equipment management;

At the proposal of the Director of the Department of Medical Equipment and Construction;

 

DECIDES:

 

Article 1. Scope of regulation

To promulgate together with this Decision the ASEAN common submission dossier template, applicable to the dossiers for circulation registration numbers for type-B, -C and -D medical devices.

Article 2. Effect

This Decision takes effect on the date of promulgation.

Article 3. Implementation organization

Chief of the Ministry Office, Chief of the Ministry Inspector, Directors and Director Generals of the Ministry of Health, Directors of Health Departments of provinces and centrally-run cities, and relevant agencies, organizations and individuals shall implement this Decision.

Any difficulties arising in the course of implementation should be promptly reported to the Ministry of Health for consideration and settlement./.

 

 

FOR THE MINISTER

THE DEPUTY MINISTER

 

Truong Quoc Cuong

 

 

 

ASEAN COMMON SUBMISSION DOSSIER TEMPLATE

(Attached to the Minister of Health’s Decision No. 2426/QD-BYT dated May 15, 2021)

 

ASEAN COMMON SUBMISSION DOSSIER

 

Name of the facility registering the circulation of medical device (hereinafter referred to as registrant) (name and address)

The registrant undertakes that the following statements are true and that it takes full responsibility before the law for them:

1. General summary

1.1. Brief and general description of medical device

General description of medical device includes: descriptive information, introduction to the medical device, purposes and indications for use, new features if any (for example, whether nano technology or artificial intelligence is used, etc.)

1.2. Commercial marketing history

Provision of a list of countries where the device has been marketed, along with the year (if any) of starting selling in that market.

1.3. Intended use and indications

Intended use and indications for use of the medical device shown on the label or instructions for that medical device.

1.4. Licensing information for circulation in different countries

Provision of information on device licensing status in the following countries: EU member countries, Japan, Canada, Australia (TGA), USA (FDA), UK, Switzerland including licensing status (approved, awaiting approval, refused, not registered for circulation, etc.), intended use, indications, date of first licensing1.

This information may be provided in a tabular format as given below:

No.

Country name or Licensing agency name

Intended use/indications

Licensing status

Licensing date

1

 

 

 

 

2

 

 

 

 

3

 

 

 

 

 

Provision of copies (or links for online search) of documents for licensing the circulation on the above-mentioned countries (if any).

1.5. Important safety or performance information of the medical device

Provision of summary information about adverse events that have occurred and corrective actions to ensure safety in the market since the medical device was introduced to the market or in the last 5 years.

Summary information about adverse events that have occurred should include, at a minimum, a description of the adverse event, the number of adverse events, or the frequency of occurrence (i.e., the total number of adverse events recorded in total number of medical devices sold). This information may be provided in a tabular format as given below:

No.

Description of adverse event

Quantity or frequency of occurrence

1

 

 

2

 

 

3

 

 

 

Information about corrective actions to ensure safety in the market, at least including date of occurrence, description of the event, name of the country or territory taking the corrective action. This information may be provided in a tabular format as given below:

No.

Date

Brief description of the event

Name of the country or territory

1

 

 

 

2

 

 

 

3

 

 

 

 

If there have been no adverse events or corrective actions since the medical device circulation or in the last 5 years, a written confirmation from the device owner is required.

Specify whether the medical device contains one of the following ingredients:

• Human or animal cells, tissues or their derivatives are used in non-living form, for example artificial heart valves originating from pigs, catgut, etc.

• Cells, tissues or derivatives of microbial or recombinant origin, for example skin plumping devices based on hyaluronic acid obtained by bacterial fermentation.

• Containing a radiation component, ionizing (e.g. X-ray), or non-ionizing (e.g. laser, ultrasound, etc.).

2. Table of conformity to the essential principles (may be in Vietnamese or English)

The table of conformity to the essential principles of medical device safety and performance issued by the medical device owner may be provided as described in the attached Appendix to demonstrate the device in compliance with relevant essential principles.

Where medical devices have been licensed for circulation in EU member states, a table of conformity with essential principles according to EU regulations can be provided.

3. Medical device description

3.1. Medical device description and features

Providing a more detailed description of the medical device attributes to explain how the device functions, the basic scientific concepts that form the fundamentals for the medical device, the component materials and accessories used in its principles of operation as well as packaging. A complete description of each functional component, material or ingredient of the medical device should be provided, with labeled pictorial representation of the device in the form of diagrams, photographs or drawings, as appropriate.

3.2. Intended use

The use for which the medical device is intended, for which it is suited according to the data supplied by the medical device owner in the instructions as well as the functional capability of the medical device.

3.3. Indications

A general description of the disease or condition that the medical device will diagnose, treat, prevent, cure or mitigate and includes a description of the target patient population for which the medical device is intended.

3.4. Instructions for use

All necessary information from the device owner, including the procedures, methods, frequency, duration, quantity and preparation to be followed for safe use of the medical device.

3.5. Contraindications

Information about cases where the medical device should not be used for patient safety reasons, for example:


 due to medical history, physiological characteristics of the patient, etc. in accordance with the information written on the label or instructions of the medical device.

This is a general description of the disease or condition and the patient population for which the medical device should not be used for the purpose of diagnosing, treating, curing or mitigating. Contraindications are conditions under which the medical device should not be used because the risk of use clearly outweighs any possible benefit.

3.6. Warnings

This is the specific hazard alert information that a user needs to know before using the medical device.

3.7. Precautions

This alerts the user to exercise special care necessary for the safe and effective use of the medical device. They may include actions to be taken to avoid effects on patients/users that may not be potentially life-threatening or result in serious injury, but about which the user should be aware. Precautions may also alert the user to adverse effects on the medical device of use or misuse and the care necessary to avoid such effects.

3.8. Potential adverse effects

These are potential undesirable and serious outcomes (death, injury, or serious adverse events) to the patient/user, or side effects from the use of the medical device, under normal conditions.

3.9. Alternative therapy

This is a description of any alternative practices or procedures for diagnosing, treating, curing or mitigating the disease or condition for which the medical device is intended.

3.10. Materials (may be in Vietnamese or English)

A description of the materials of the medical device and their physical properties to the extent necessary to demonstrate conformity with the relevant Essential Principles.

For a medical device other than in vitro diagnostic medical devices:

- List of materials of the medical device making either direct (e.g. with the mucous membrance) and or indirect (e.g. during extracorporeal circulation of body fluids) with the human body, enclosed their chemical, biological and physical properties.

- For medical device intended to emit ionizing radiation, information on the radiation source (e.g. radioisotopes) and the material used for shielding of unintended radiation from users and patients shall be provided.

- Where there are specific concerns related to the safety of materials used in the medical device, e.g. impurities, residue levels and exposure to plasticizers such as Bis(2-ethylhexyl) phthalate (DEHP), it is necessary to additionally provide a certificate of raw material in conformity with relevant standards, analysis certificate, or risk assessment on the safety of the materials used. Depending on the risk of exposure to these materials, additional measures may be required, such as notifying the user of the presence of these materials by labeling the device.

For in vitro diagnostic reagents, calibrators and control materials:

- Provide the list of materials used for manufacture of the medical device, including the material name, role in finished product.

- Provide information about the composition, biological properties, and origin of the materials involved in the reaction for testing: antigens, antibodies, enzymes, conjugates, PCR primers, probes, calibrators, controls, etc.

- For medical device containing substances subject to special control (narcotics or radiation-generating materials, non- ionizing or ionizing), full information on ingredients, content and role of these substances, for example Buprenorphine in the drug testing kit, Iodide-131 in the Radioimmunoassay kit, radio-labeled Phosphorus-32 DNA probes in Southern blots), etc. must be provided.

- For in vitro diagnostic medical device of types C, D: information about standards or test reports of materials participating in the reaction (except stabilizers) must be provided. The information is presented for each active material (antigen, antibody, conjugate, etc.).

3.11. Other relevant specifications

The functional characteristics and technical performance specifications for the medical device including, as relevant, accuracy, sensitivity, specificity of measuring and diagnostic medical devices, reliability and other factors; and other specifications including chemical, physical, electrical, mechanical, biological, software, sterility, stability, storage and transport, and packaging to the extent necessary to demonstrate conformity with the relevant Essential Principles.

3.12. Other information

Other important descriptive characteristics not detailed above, to the extent necessary to demonstrate conformity with the relevant Essential Principles (for example, the biocompatibility category for the medical device, etc.).

Note: In case the above information is stated in the instructions, reference may be made to the instructions.

4. Summary of design verification and validation documents (may be in Vietnamese or English)

This section should summarize or reference or contain design verification and design validation data to the extent appropriate to the complexity and risk class of the medical device.

Such documentation should typically include:

• Declarations or certificates of conformity to the “recognized” standards listed as applied by the product owner; and/or

• Summaries or reports of tests and evaluations based on other standards, manufacturer methods and tests, or alternative ways of demonstrating compliance.

For example: The completed table of conformity to the essential principles mentions that manufacturer uses a recognized standard was to demonstrate conformity to one Essential Principle, a declaration of conformity to the standard, or a certification of conformity, and a summary of the test data, if the standard does not include performance requirements, must be provided.

The data summaries or tests reports and evaluations would typically cover, as appropriate to the complexity and risk class of the medical device:

- A listing of and conclusions drawn from published reports that concern the safety and performance of aspects of the medical device with reference to the Essential Principles;

- Engineering tests;

- Laboratory tests;

- Biocompatibility tests;

- Animal tests;

- Simulated use;

- Software validation.

4.1. Pre-clinical studies

Details must be provided on all biocompatibility tests conducted on materials used in a medical device All materials that are significantly different must be characterized. Information describing the tests, the results and the analyses of data must be presented.

Complete pre-clinical physical test data must be provided, as appropriate. The report must include the objectives, methodology, results and product owner’s conclusions of all physical studies of the medical device and its components. Physical testing must be conducted to predict the adequacy of medical device response to physiological stresses, undesirable conditions and forces, long-term use and all known and possible failure modes.

Pre-clinical animal studies used to support the probability of effectiveness in humans must be reported. These studies must be undertaken using good laboratory practices. The objectives, methodology, results, analysis and product owner’s conclusions must be presented. The study conclusion should address the medical device’s interactions with animal fluids and tissues and the functional performance of the medical device in the experimental animal model(s). The rationale (and limitations) of selecting the particular animal model should be discussed.

Evidence of electrical safety and electromagnetic compatibility should be provided. For example, if the owner declares that the product meets the requirements of IEC 60601-1 and IEC 60601-1-2, a summary test report and/or certificate of conformity must be provided to demonstrate that the device meets these standards.

For sterilized medical device, sterilization appraisal reports must be provided. If the sterilant is toxic or produces toxic residuals (e.g. ethylene oxide residuals), test data and methods that demonstrate that post-process sterilant or residuals are within the acceptable limits must be presented.

For net-work connected, Internet-connected and wireless enable medical devices, evidence supporting the cyber security of such devices shall be provided. For example: analyzing risks and possibilities of cyber-attacks, cyber security control methods, plans, processes or mechanisms to monitor, promptly detect and manage security-related threats during the shelf life of the medical device. A declaration from the manufacturer or owner stating that the medical device is connected to the network, Internet or it is a wireless enable device, ensuring cyber security.

4.1.1. Software verification and validation studies (if applicable)

The correctness of a software product is another critical product characteristic that cannot be fully verified in a finished product. The product owner must provide evidence that validates the software design and development process. This information should include the results of all verification, validation and testing performed in-house and in a user’s environment prior to final release, for all of the different hardware configurations identified in the labeling or the instructions, as well as representative data generated from both testing environments.

4.1.2. Medical devices containing biological materials

Results of studies substantiating the adequacy of the measures taken with regards to the risks associated with transmissible agents must be provided. This will include viral clearance results for known hazards. Donor screening concerns must be fully addressed and methods of harvesting must also be fully described. Process validation results are required to substantiate that manufacturing procedures are in place to minimize biological risks.

4.2. Pre-clinical studies for in vitro diagnostic reagents, calibrators and control materials

4.2.1. Analytical performance

Provision of research reports on the analytical performance of the medical device, including analytical sensitivity, analytical specificity, limit of detection (LOD), limit of quantification (LOQ), linearity, detection range, precision, repeatability, influencing factors, durability, etc. Performance criteria will depend on each medical device.

Information about samples used for evaluation (negative and positive control samples, baseline standard samples, standard panels, influencing substances, etc.).

Research reports should include the objectives, methods, results, and conclusions of the research. The results and conclusions must clearly demonstrate that the product has characteristics suitable for its intended use.

4.2.2. Stability

Provision of stability study reports, including real-time stability and stability under accelerated aging conditions (as appropriate). In cases where real- time stability study is not performed but only under accelerated aging conditions, a full and reasonable explanation is required.

Provision of in-use stability study reports for products that are used multiple times after opening.

Stability study reports during transportation, performed under real or simulated conditions shall be provided.

Stability study reports should include the objectives, methods, results, and conclusions.

4.3. Clinical evidence

A clinical evaluation report of the medical device should be provided. Where applicable, this evaluation may take the form of a systematic review of existing bibliography, clinical experience with the same or similar medical devices, or by clinical investigation. Clinical investigation is most likely to be needed for higher risk class medical devices, or for medical devices where there is little or no clinical experience.

The clinical evaluation report should include the purpose and context of the clinical evaluation, clinical input data, data evaluation and analysis, and conclusions regarding the safety and performance of the medical device.

The clinical evaluation report should have all the necessary information as an independent document for review by regulatory agencies. This should summarize:

- The technology used by that medical device, indications, statements about the safety and clinical performance of that medical device, if any.

- Nature, scope and scale of clinical data assessed.

- Clinical data and recognized standards proving the safety and performance of the medical device.

4.3.1. Use of existing bibliography

Copies are required of all literature studies, or existing bibliography, that the product owner is using to support safety and effectiveness.

Clinical evidence of effectiveness may comprise medical device-related investigations conducted domestically or other countries. It may be derived from relevant publications in international medical journals. The documented evidence submitted should include the objectives, methodology and results presented in context, clearly and meaningfully. Before drawing conclusions about the results of clinical investigation, there must be a discussion in context with the published literature.

4.3.2. Use of clinical experience

Clinical experience refers to clinical data obtained from clinical use of the product, not from clinical investigations. Clinical experience may be with the same product or a similar product.

Clinical experience may be obtained from the following data:

- Post-marketing surveillance reports from product owners, regulatory agencies, cohort studies (may contain unpublished long-term study data on safety and effectiveness).

- Data on adverse events that have occurred, from product owners or regulatory agencies.

- Data from patients using the medical device in the aid program before marketing the product.

- Information on clinically relevant corrective actions such as recalls, notifications, hazard warnings.

4.3.3. Use of clinical investigations

A clinical investigation is a systematic investigation or study in or on one or more human subjects, undertaken to assess the safety and performance of a medical device.

A clinical investigation may be conducted by the medical device owner or by a third party on behalf of the owner. Clinical investigations should be designed, performed and reported in accordance with ISO 14155, Parts 1 and 2, Clinical Investigations of Medical Devices for Human Subjects, or in accordance with an equivalent standard, and in compliance with local regulations.

Clinical investigations shall be carried out in accordance with the ethical principles laid down in the Helsinki Declaration.

For in vitro diagnostic medical devices: Clinical investigation is performed to establish or confirm the clinical performance of in vitro diagnostic medical devices. The manufacturer must have clinical evidence supporting its clinical claims, including diagnostic sensitivity (clinical sensitivity) and diagnostic specificity (clinical specificity).

5. Medical device labeling

This is the descriptive and informational product literature that accompanies the medical device any time while it is held for sale or shipped, including instructions, pack labeling. This section should summarize or reference or contain the following labeling data to the extent appropriate to the complexity and risk class of the medical device, which is generally considered as “labeling”:

• Labels on the medical device and its packaging;

• Instructions for use;

• Any information and instructions given to the patient, including instructions for any procedure the patient is expected to perform (if applicable).

5.1. Samples of labels on the medical device and its packaging

This is the printed, written or graphic product information provided on or attached to one or more levels of packaging, including the outer packaging or the outside container wrapper. If it is physically impossible to include samples of labels (e.g. large warning labels affixed onto an X-ray machine), alternative submission methods (e.g. photographs or technical drawings), to the extent appropriate, will suffice to meet the requirements of this section.

The samples of labels must meet requirements in accordance with the applicable laws. For an imported medical device, the label sample must include both the original label and the supplementary label in Vietnamese language.

The label sample must include all products in the circulation registration dossier. In case the dossier is submitted according to the medical device family, a representative label sample can be submitted but must note the differences between the label samples of those products.

5.2. Instructions for use in Vietnamese language

The instructions for use contain directions under which the physician or end-user can use a medical device safely and for its intended purpose. This should include information on indications, contraindications, warnings, precautions, potential adverse effects, alternative therapy and the conditions that should be managed during normal use to maintain the safety and effectiveness of the medical device.

For in vitro diagnostic medical devices, the instructions for use should contain the following information:

- Type and code (if any) of the in vitro diagnostic medical device.

- Intended use:

• To detect or measure something;

• Medical device functions (e.g. screening, monitoring, diagnosis or supporting diagnosis, prognosis, prediction);

• Use the same medical device (testing machine) whether automated or non-automated;

• Quantitative or semi-quantitative or qualitative or nominal, etc.;

• Type of sample used in the test (e.g. serum, plasma, whole blood, tissue biopsy, urine);

• Test population.

- Instructional information: product used in in vitro diagnostics.

- Users (e.g. laypersons, medical staff, etc.).

- Principle of testing.

- Description of reagents, calibrators, standards, control materials and limitations on their use (e.g. only suitable for use with a specific testing machine).

- List of materials provided and list of materials specifically requested but not provided.

- For in vitro diagnostic medical device used in combination with other medical devices and/or in combination with non-medical devices:

• Information to identify these devices or equipment, including important performance characteristics.

• Information on known limitations associated with that device or equipment.

- Special storage conditions (e.g. temperature, light, humidity), conditions when using the medical device, if relevant.

- Stability during use, which may include storage conditions and shelf life after first opening; storage conditions and stability of working solutions, as appropriate. The instructions for use must clearly state the product's lifespan.

- If the medical device is provided as sterile, it is necessary to clearly state the medical device 's sterility, sterilization method, and instructions for handling in case the sterile packaging is damaged before use.

- Information to enable the user to be informed of any warnings, precautions, measures to be taken and limitations on use of the in vitro diagnostic medical device. This information should include, as appropriate:

• Warnings, precautions and/or measures to be taken in the case of in vitro diagnostic medical device damage or deterioration reflected through a change in the appearance of the medical device that may affect performance;

• Warnings, precautions and/or measures to be taken regarding exposure to foreseeable external influences or environmental conditions, such as external magnetic, electrical and electromagnetic fields, electrical discharge, radiation related to diagnostic or therapeutic procedures, pressure, humidity or temperature;

• Warnings, precautions and/or measures to be taken regarding possible risks during the use of the medical device for specific diagnosis, evaluation, treatment (e.g. electromagnetic interference emitted by the device, affecting other devices);

• Precautions regarding materials incorporated into the medical device that contain carcinogenic, mutagenic or reproductively harmful substances, endocrine disruptors, or other substance that may cause sensitivity or allergy in the patient or user.

- Conditions for collecting, manipulating and preparing samples

- Instructions for preparing or handling the medical device before use, such as sterilization, assembly, calibration, etc., so that the device is used as intended by the owner.

- Information needed to determine when the in vitro diagnostic medical device is properly installed and ready to perform safely and as intended by the owner, along with the following information, when relevant:

• Details of the nature and frequency of preventive and routine maintenance, including cleaning or disinfection;

• Consumables and instructions to replace them;

• Information about the calibrations that need to be performed to ensure the medical device operates properly and safely throughout its life cycle;

• Measures to reduce the risk to persons involved in the installation, calibration or repair of, for example, contaminated surfaces.

- Recommendations on quality control procedures, if necessary.

- Instructions on how to retrieve the values assigned to calibrators and control materials, including information to identify reference materials and reference measurement procedures used.

- Test procedure, including calculation and interpretation of results, and recommendation of additional confirmatory testing if appropriate.

- Analytical performance characteristics, such as sensitivity, specificity and precision (which is a combination of accuracy and precision).

- Clinical performance characteristics, such as diagnostic sensitivity and diagnostic specificity, if relevant.

- Reference intervals, if relevant.

- Information on interferences or limitations (e.g. visual evidence of hyperlipidemia or hemolysis, specimen storage duration) that may affect the performance of the analysis.

- Warnings or precautions to be taken regarding disposal of the device, accessories and consumables used together, if any. This information should include, as appropriate:

• Bacterial or microbial contamination hazards (e.g. consumable materials contaminated with potentially infectious substances of human origin);

• Environmental hazards (e.g. batteries or materials that emit potentially dangerous levels of radiation);

• Physical hazards (e.g. explosion).

- For in vitro diagnostic medical device used by layperson, users should be advised to consult a health care professional before making any medical decisions.

- References, if relevant.

- Name and address of the owner in a format that is identifiable and allows identification of information about the owner. In case the instructions for use contain information about the manufacturing facility and origin, this information must include complete information about the name and address of the manufacturer.

- Document identification information, such as version number or release date.

6. Risk analysis (may be in Vietnamese or English)

This section should summarize or reference or contain the results of the risk analysis. This risk analysis should be based upon international or other recognized standards, and be appropriate to the complexity and risk class of the medical device.

6.1. Results of risk analysis

A list of possible hazards for these medical devices shall be provided. Indirect risks from medical devices may result from medical device-associated hazards, such as moving parts, which lead to sustained injury, or from user-related hazards, such as ionizing radiation from an X-ray machine. The evaluation of these risks against the claimed benefits of the medical device and the method(s) used to reduce risk to acceptable levels must be described. The individual or organization that carries out the risk analysis must be clearly identified. The technique used to analyze risk must be specified, to ensure that it is appropriate for the medical device and the risk involved.

7. Manufacturing information (may be in Vietnamese or English)

This section should summarize or reference or contain documentation related to the manufacturing processes, including quality assurance measures, which is appropriate to the complexity and risk class of the medical device.

7.1. Manufacturer information

List the names and addresses of all manufacturers involved in the manufacturing and sterilization process (including third-party manufacturers and sterilizers).

7.2. Manufacturing process

Information on the manufacturing process shall be provided in sufficient detail to allow a general understanding of the manufacturing processes. Detailed proprietary information on the manufacturing process is not required. The information may be presented in the form of a process flow chart showing an overview of production, controls, assembly, in-process and final product testing, and packaging of the finished product.

If multiple manufacturers are involved in the manufacturing process to complete a product, it is necessary to clearly state which activities each manufacturer is involved in.

 

 

Legal representative

Signature (Full name and title)

Certification by stamp or digital signature

 

_________________

1 For products licensed for circulation in EU member countries, if there is no information on the date of first licensing, the most recent licensing date may be recorded.

 
 
 

APPENDIX

TABLE OF CONFORMITY TO THE ESSENTIAL PRINCIPLES (may be in Vietnamese or English)

 

Name of the medical device owner:

Medical device(s):

Product code(s) (if any):

Essential principle

Applicable to the device? (Yes/No)

Applicable standard

Evidence demonstrating the conformity

General requirements

1. Medical devices should be designed and manufactured in such a way that, when used under the conditions and for the purposes intended and, by virtue of the knowledge and experience of intended users, they will not compromise the clinical condition or the safety of patients, or the safety and health of users or other persons, provided that any risks which may be associated with their use constitute acceptable risks when weighed against the benefits to the patient and are compatible with a high level of protection of health and safety.

2. The solutions adopted by the medical device owner for the design and manufacture of the devices shall conform to safety principles, taking account of the scientific and technical development. In selecting the most appropriate solutions for the design and manufacture of the medical device, to minimize the risks associated with the use of medical device, the product owner should apply the following principles:

• identify hazards and the associated risks arising from the intended use and foreseeable misuse,

• eliminate or reduce risks as far as possible (inherently safe design and construction),

• where appropriate take adequate protection measures including alarms if necessary, in relation to risks that cannot be eliminated, and

• inform users of any residual risks.

3. The medical device must achieve the performance intended by the medical device owner and be designed, manufactured, and packaged to suit one or more functions within the scope of the definition of a medical device.

4. The technical characteristics and properties in Clauses 1, 2 and 3 shall not have an adverse effect to the extent that the health or safety of the patient or user and of others is impaired during throughout the life cycle of the medical device as expected by the medical device owner, when the medical device is subject to impacts that may occur under normal use conditions and has been maintained, calibrated according to the instructions of the medical device owner.

5. Medical devices must be designed, manufactured and packaged so that their technical characteristics and properties when used for their intended purpose will not be adversely affected during transport and storage if the transport and storage are carried out in accordance with instructions and information provided by the medical device owner.

6. The benefits must outweigh any unwanted side effects to the intended effect.

7. Medical devices require clinical evidence, appropriate to their use and risk classification, to demonstrate that the medical device complies with the correspondingly design principles. Clinical evaluation must be performed.

Design and manufacturing requirements

8. Chemical, physical and biological properties

8.1. Medical devices must be designed and manufactured to ensure that they meet the requirements for technical characteristics and properties set out in Clauses 1 thru 6 of “General Requirements”. Attention should be paid to:

• the selection of materials used, specifically regarding toxicity, flammability,

• physical and chemical properties of the materials used,

• Compatibility between the materials used and biological tissues, cells, body fluids, and specimens, taking into account the intended purpose of the medical device,

• The selection of materials used must reflect issues such as abrasion, hardness and resistance of the metal.

8.2. Medical devices must be designed, manufactured and packaged to minimize the risk posed by contaminants and residues to patients and those involved in transportation, storage and use of the medical device. In minimizing risk, special attention should be paid to the duration and frequency of tissue exposure during the process of transporting, preserving and using the medical device.

8.3. Medical devices must be designed and manufactured so that they can be used safely with the materials, substances and gases to which they are exposed under normal use or in routine process. If medical devices are used to administer drugs, they must be designed and manufactured so that they are compatible with relevant medicinal products and that the action of the drug is maintained for its intended use.

8.4. Where a medical device contains a substance that, if used separately, could be considered a medicinal product under relevant laws and has the effect of supporting the effects of the medical device on the body, the safety, quality and performance of that medical device as well as the safety, quality and effectiveness of the combined medicinal product must be verified. “Medicinal product” referred to herein includes stable derivatives from human blood or plasma.

8.5. Medical devices must be designed and manufactured to minimize risks caused by substances that may leach or leak from the medical device.

8.6. Medical devices must be designed and manufactured to minimize risks caused by inadvertent penetration or release of substances from the medical device, taking into account the nature of the environment in which the medical device is intended to be used.

9. Bacterial infection and microbial contamination

9.1. Medical devices and manufacturing processes should be designed to eliminate or reduce, as far as reasonably and appropriately possible, the risk of infection to any person. A design must:

• be easy to operate, and, when necessary:

• minimize, as far as reasonably and appropriately possible, microbial leakage and/or microbial exposure during use;

• minimize the spread of infection from patients, users or other people to the medical device or test samples and vice versa during use.

9.2. In case a medical device is composed of substances of biological origin, the risk of contamination must be minimized to the maximum reasonable and appropriate level by selecting the source, sample donor and appropriate substances, and using inactivation, conservation, testing and control procedures, as appropriate. This requirement may not apply to certain in vitro diagnostic medical devices if viral activity and infectious agents are an essential part of the intended use of the product or the removal or inactivation will affect the effectiveness of that in vitro diagnostic medical device.

9.3. Products made from non-living tissues, cells, and substances of animal origin that are determined to be medical devices shall be considered ingredients of animal origin and will apply relevant regulations and subject to veterinary control and supervision for the intended use of this tissue. Medical device owners need to maintain information about the geographical origin of animals. The processing, storage, inspection and handling of tissues, cells and substances of animal origin must be carried out to ensure optimal safety. In particular, safety with regard to viruses and other infectious agents must be achieved by inactivation or removal methods appraised during manufacturing. This requirement may not apply to certain in vitro diagnostic medical devices if other viral activity and infectious agents are an essential part of the intended use of the product or the removal or inactivation will affect the effectiveness of that in vitro diagnostic medical device.

9.4. For products made from cells, tissues and derivatives of bacterial or recombinant origin that are determined to be medical devices, the selection of sample sources/donors, processing and preservation, testing and handling of cells, tissues and derivatives of such origin must be carried out to achieve optimal safety level. In particular, safety with regard to viruses and other infectious agents must be handled by performance of inactivation or removal methods appraised during manufacturing. This requirement may not apply to certain in vitro diagnostic medical devices if other viral activity and infectious agents are an essential part of the intended use of the product or the removal or inactivation will affect the effectiveness of that in vitro diagnostic medical device.

9.5. For products composed of non-living tissues, cells, and substances of human origin determined to be in vitro diagnostic medical devices, the selection of sample sources, donors, and/or substances of human origin, the processing, storage, testing and handling of tissues, cells and substances of such origin must be carried out to ensure optimal safety. In particular, safety with regard to viruses and other infectious agents must be achieved by inactivation or removal methods appraised during manufacturing. This requirement may not apply to certain in vitro diagnostic medical devices if other viral activity and infectious agents are an essential part of the intended use of the product or the removal or inactivation will affect the effectiveness of that in vitro diagnostic medical device.

9.6. Medical devices in a special microbiological state must be designed, manufactured and packaged to ensure that the products retain their properties when placed on the market and during transportation and preservation according to the regulations of the medical device owners.

9.7. Sterile medical devices must be designed, manufactured and packaged to ensure that the products remain sterile when placed on the market and during transportation and preservation according to the regulations of the medical device owner.

9.8. Medical devices that are sterile or in a special microbiological state must be handled, manufactured, and, if necessary, sterilized using appropriate, validated methods.

9.9. Medical devices intended to be sterilized must be manufactured under suitably controlled (e.g. environmental) conditions.

9.10. Packagings for non-sterile medical devices must be capable of keeping the products at the prescribed level of cleanliness. If the medical device needs to be sterilized before use, it must be minimized to maximum risk of microbial infection; packaging must be appropriate, taking into account the sterilization method prescribed by the medical device owner. The medical device must be manufactured under appropriately controlled conditions.

9.11. The packagings and/or labels of medical devices must be distinguishable from identical or similar products on the market in

both sterile and non-sterile conditions.

10. Manufacturing and environmental characteristics

10.1. If the medical device is intended to be used in combination with other medical devices or equipment, the entire combination, including the connection system, must be safe without affecting the determined performance of the medical device or equipment used with it. Any limitations on use in such combination must be stated on the label and/or in the instructions.

10.2. Medical devices must be designed and manufactured to eliminate or minimize to a reasonable and appropriate level:

• risk of injury, related to their physical features, including, as appropriate, mass/pressure ratio, dimensional and design characteristics;

• risks related to environmental conditions or reasonably foreseeable external influences, such as magnetic fields, external electrical and electromagnetic effects, electrostatic discharge, pressure, humidity, temperature or pressure changes and acceleration;

• risks associated with the use of medical devices in combination with materials, substances or gases to which they may come into contact under normal conditions of use;

• risks due to accidental penetration of substances into the medical device;

• risks due to not correctly identifying test samples;

• risks due to interaction with other medical devices often used in research or for certain treatments;

• risks arising from failure to maintain or calibrate (as with implantable devices), from aging of the materials used or from the loss of accuracy of some measurement or control mechanism.

10.3. Medical devices should be designed and manufactured in such a way as to remove or appropriately reduce the risks of fire or explosion during normal use and in single fault condition. Particular attention should be paid to medical devices whose intended use includes exposure to or in association with flammable or explosive substances or substances which could cause combustion.

10.4. Medical devices should be designed and manufactured in such a way as to facilitate the safe disposal of waste substances.

11. Medical devices with a diagnostic or measuring function

11.1. Medical devices with a measuring function should be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose. The limits of accuracy, precision and stability should be indicated by the medical device owner.

11.2. Medical devices should be designed and manufactured in such a way as to provide sufficient accuracy, precision and stability for their intended purpose, based on appropriate scientific and technical methods. Specifically, the design must address sensitivity, specificity, precision, repeatability, reproducibility, control of known confounders, and limits of detection, if any.

11.3. Where the performance of medical devices depends on the use of calibrators or control materials, the traceability of values assigned to such calibrators and/or control materials should be ensured through a quality control system.

11.4. Any measurement, monitoring or display scale functions of medical devices should be designed in line with design science principles, taking account of the intended purpose of medical devices.

11.5. When values are expressed in numbers, if possible, standard units, generally accepted, and clearly understood by the user of the medical device should be used.

12. Protection against radiation

12.1. General requirements

12.1.1. Medical devices should be designed, manufactured and packaged in such a way that exposure of patients, users, other persons, to radiation is appropriately reduced in a manner that is compatible with the intended purpose, whilst not restricting the application of appropriate specified levels for diagnostic and therapeutic purposes.

12.2. Intended radiation

12.2.1. Where medical devices are intended to emit visible and/or invisible hazardous, or potentially hazardous, radiation for a specific medical purpose, the benefit is considered is greater than the inherent risks of such radiation, and the radiation can be controlled by the user, such medical devices should be designed and manufactured to ensure reproducibility of relevant variable parameters within an acceptable tolerance.

12.2.2. Where medical devices are intended to emit visible and/or invisible hazardous, or potentially hazardous, radiation, they should be fitted with visual displays and/or audible warnings of such emissions.

12.3. Unintentional radiation

12.3.1. Medical devices should be designed and manufactured in such a way that the exposure of users, other persons, or patients, to the emission of unintended, stray or scattered radiation is appropriately reduced.

12.4. Operating instructions

12.4.1. The operating instructions for medical devices emitting radiation should contain detailed information as to the nature of the emitted radiation, the means of protecting the users, other persons, or patients, and ways of avoiding misuse and of appropriately reducing the risks inherent to installation.

12.5. Ionizing radiation

12.5.1. Medical devices emitting ionizing radiation should be designed and manufactured in such a way as to ensure that the quantity, shape and distribution of equipment (or quality) of the emitted radiation can be varied and controlled where possible, taking into account the intended use.

12.5.2. Medical devices emitting ionizing radiation intended for medical imaging should be designed and manufactured in such a way as to achieve an image and/or output quality that are appropriate to the intended medical purpose whilst minimizing radiation exposure of the patient and user.

12.5.3. Medical devices emitting ionizing radiation intended for medical imaging should be designed to provide good monitoring and control of dose used, beam type, energy and, when possible, monitoring and control of the energy distribution of the radiation beam.

13. Requirements for medical device connected or equipped with a power source.

13.1. Medical devices that incorporate electronic programmable systems, including software, should be designed to ensure repeatability, reliability and performance in line with their intended use. In the event of a single fault condition (SFC), appropriate means should be adopted to eliminate or appropriately reduce consequent risks or impairment of performance.

13.2. For medical devices that incorporate software or are software as a medical device, the software should be evaluated based on advanced scientific and technical foundations, taking into account the principles of development life cycle, risk management, verification and validation.

13.3. Medical devices where the safety of the patient depends on an


internal power supply should be equipped with a means of determining the state of the power supply.

13.4. Medical devices where the safety of the patient depends on an external power supply should include an alarm system to signal any power failure.

13.5. Medical devices intended to monitor one or more clinical parameters of a patient should be equipped with appropriate alarm systems to alert the user of situations which could lead to death or severe deterioration of the patient's state of health.

13.6. Medical devices should be designed and manufactured in such a way as to appropriately reduce the risks of creating electromagnetic interference which could impair the operation of such devices or any devices located in close proximity to such devices.

13.7. Medical devices should be designed and manufactured in such a way as to provide a level of intrinsic immunity to electromagnetic interference such that is adequate to enable them to operate as intended.

13.8. Protect against electrical risks

13.8.1. Medical devices should be designed and manufactured in such a way as to appropriately reduce the risk of accidental electric shocks to the user or any other person, both during normal use of the medical device and in the event of a single fault condition (SFC) in the medical device, provided the medical device is installed and maintained as indicated by the product owner.

14. Protect against mechanical risks

14.1. Medical devices should be designed and manufactured in such a way as to protect patients and users against mechanical risks related to the use of devices.

14.2. Medical devices should be designed and manufactured in such a way as to appropriately reduce the risks arising from vibration generated by the medical devices, taking account of technical progress and of the means available for limiting vibrations, particularly at source, unless the vibrations are part of the specified performance.

14.3. Medical devices should be designed and manufactured in such a way as to appropriately reduce the risks arising from the noise emitted, taking account of technical progress and of the means available to reduce noise, particularly at source, unless the noise emitted is part of the specified performance.

14.4. Terminals and connections to electrical, gas or hydraulic and pneumatic power supplies controlled by the user should be designed and constructed in such a way as to minimize any possible risks.

14.5. Accessible parts of medical devices (excluding the parts or areas intended to supply heat or reach given temperatures) and their surroundings should not attain potentially dangerous temperatures under normal use.

15. Protect against risks posed to the patient by the supply of energy or substances

15.1. Medical devices for supplying the patient with energy or substances should be designed and manufactured in such a way that the speed and/or amount to be delivered can be set and maintained accurately enough to ensure the safety of the patient and user.

15.2. Medical devices should be fitted with the means of preventing and/or indicating any inadequacies in the speed and/or amount of energy delivered or substances delivered which could pose a danger. Devices should incorporate suitable means to appropriately reduce the risk of accidental release of dangerous levels of energy or substances.

15.3. Monitoring and display scale functions must be clearly identified on medical device. Where a medical device has instructions necessary for its operation or displays operating or adjustment parameters by visual means, such information must be understandable to the user, and to the patient if needed.

16. Active implantable medical devices

16.1. Active implantable medical devices must have clear information to identify:

• Type of medical devices;

• Medical device owner; and

• Manufacturing year of medical devices.

16.2. That information must be readable without the need for surgery by the implant recipient.

17. Protect against risks to patients posed by medical devices intended by the manufacturer for use by lay users

17.1. Medical devices for use by lay users should be designed and manufactured in such a way that they perform appropriately for their intended use/purpose taking into account the skills and the means available to lay users and the influence resulting from variation that can be reasonably anticipated in the lay user's technique and environment. The information and instructions provided by the medical device owner should be easy for the lay user to understand and apply.

17.2. Medical devices for use by lay users should be designed and manufactured in such a way as to appropriately reduce the risk of error by the intended user in the handling of the medical device and, if applicable, in the interpretation of the results.

17.3. Medical devices for use by lay users should, where appropriate, include means by which the lay user can verify that, at the time of use, the medical device will perform as intended by the medical device owner.

18. Information provided by the medical device owner

18.1. Users must be provided with the necessary information to identify the medical device, determine the owner of the medical device, and explain information to use the medical device safely, etc.

19. Clinical investigations

19.1. Clinical investigations on human subjects shall be carried out in accordance with the spirit of the Helsinki Declaration. This includes every step in the clinical investigation from first consideration of the need and justification of the study to publication of the results.

If the Essential Principles are applicable to the device, then the answer is “Yes”;

 If the answer is “No”, then the Essential Principles are inapplicable to the device;

and this should be briefly explained. For example: For a medical device that does not incorporate biological substances, the answer to Essential Principle regarding materials of biological origin given in the column “Applicable to the medical device?” would be “No”; and “The medical device does not incorporate biological substances”.

 

(The list of reference standards or internal standards and procedures applied by the manufacturer for the product to meet the Essential Principles shall be provided. For example:

- ISO 13485

- ISO 14971

- EN ISO 18113-1

- IEC 61326-1

- IEC 61326-2-6

- IEC 60601-1)

(Name the technical documents that demonstrate the product meets the essential principles, e.g. certificates, research reports, test results, etc.)


 

Please log in to a subscriber account to see the full text. Don’t have an account? Register here
Please log in to a subscriber account to see the full text. Don’t have an account? Register here
Processing, please wait...
LuatVietnam.vn is the SOLE distributor of English translations of Official Gazette published by the Vietnam News Agency

ENGLISH DOCUMENTS

LuatVietnam's translation
Decision 2426/QĐ-BYT DOC (Word)

This utility is available to subscribers only. Please log in to a subscriber account to download. Don’t have an account? Register here

Decision 2426/QĐ-BYT PDF

This utility is available to subscribers only. Please log in to a subscriber account to download. Don’t have an account? Register here

* Note: To view documents downloaded from LuatVietnam.vn, please install DOC, DOCX and PDF file readers
For further support, please call 19006192

SAME CATEGORY

loading